New Automated Periostin Immunoassay Developed

By LabMedica International staff writers
Posted on 11 Jan 2017
Periostin is being investigated as a potential biomarker for T-helper-2 (Th2)-driven asthma or eosinophilic inflammation and may help to identify patients more likely to benefit from interleukin-13-targeted treatments.

The use of serum interleukin-13 (IL-13) itself as a biomarker is challenging because it is expressed locally in inflamed tissue, resulting in very low concentrations in serum and concentrations of serum IL-13 appear to be similar between healthy controls and subjects with asthma.

Scientists at Abbott Laboratories (Abbott Park, IL, USA) and their colleagues assessed the company’s ARCHITECT Periostin Immunoassay performance in terms of precision, sensitivity, linearity, interference from classical immunoassay interferents and representatives of common asthma medications, specimen handling, and isoform reactivity. The assay has been developed for use with the ARCHITECT Immunoassay i System.

Samples from subjects with idiopathic pulmonary fibrosis, subjects with asthma used in periostin concentration assessments, and normal samples from healthy volunteers were obtained. The test is a monoclonal antibody (mAb) sandwich two-step immunoassay for the quantitative determination of periostin in human serum using chemiluminescent magnetic immunoassay technology. Periostin, the analyte, is captured by microparticles coated with an anti-periostin mAb and subsequently detected with a second anti-periostin mAb conjugated with Acridinium.

The team reported that the percentage of coefficient of variation (CVs) for five-day total precision, assessed using two instruments was less than 6% across two controls and one serum-based panel. Limit of quantitation was 4 ng/mL, using dilution adjusted concentration, suiting the needs for the application. Dilution analysis yielded linear results and no endogenous sample or drug interferences were observed. All known periostin isoforms expressed in the mature human lung were detected by the assay.

The authors concluded that their analysis of periostin concentration in serum samples appeared to show slightly elevated mean periostin concentrations in samples from subjects with uncontrolled asthma compared with apparently healthy volunteers. This is consistent with reports of elevated serum periostin concentration in subjects with asthma, particularly those with Th2-driven asthma or eosinophilic inflammation. The study was published in the January 2017 issue of the journal Clinica Chimica Acta.

Related Links:
Abbott Laboratories


Platinum Member
COVID-19 Rapid Test
OSOM COVID-19 Antigen Rapid Test
Magnetic Bead Separation Modules
MAG and HEATMAG
Complement 3 (C3) Test
GPP-100 C3 Kit
New
Gold Member
Fully Automated Cell Density/Viability Analyzer
BioProfile FAST CDV
Read the full article by registering today, it's FREE! It's Free!
Register now for FREE to LabMedica.com and get complete access to news and events that shape the world of Clinical Laboratory Medicine.
  • Free digital version edition of LabMedica International sent by email on regular basis
  • Free print version of LabMedica International magazine (available only outside USA and Canada).
  • Free and unlimited access to back issues of LabMedica International in digital format
  • Free LabMedica International Newsletter sent every week containing the latest news
  • Free breaking news sent via email
  • Free access to Events Calendar
  • Free access to LinkXpress new product services
  • REGISTRATION IS FREE AND EASY!
Click here to Register








Channels

Clinical Chemistry

view channel
Image: The 3D printed miniature ionizer is a key component of a mass spectrometer (Photo courtesy of MIT)

3D Printed Point-Of-Care Mass Spectrometer Outperforms State-Of-The-Art Models

Mass spectrometry is a precise technique for identifying the chemical components of a sample and has significant potential for monitoring chronic illness health states, such as measuring hormone levels... Read more

Hematology

view channel
Image: The CAPILLARYS 3 DBS devices have received U.S. FDA 510(k) clearance (Photo courtesy of Sebia)

Next Generation Instrument Screens for Hemoglobin Disorders in Newborns

Hemoglobinopathies, the most widespread inherited conditions globally, affect about 7% of the population as carriers, with 2.7% of newborns being born with these conditions. The spectrum of clinical manifestations... Read more

Microbiology

view channel
Image: The T-SPOT.TB test is now paired with the Auto-Pure 2400 liquid handling platform for accurate TB testing (Photo courtesy of Shutterstock)

Integrated Solution Ushers New Era of Automated Tuberculosis Testing

Tuberculosis (TB) is responsible for 1.3 million deaths every year, positioning it as one of the top killers globally due to a single infectious agent. In 2022, around 10.6 million people were diagnosed... Read more
Copyright © 2000-2024 Globetech Media. All rights reserved.