Novel Procalcitonin Immunoassay Analytically Evaluated

By LabMedica International staff writers
Posted on 25 Jan 2017
Print article
Image: The LUMIPULSE G1200 is a robust mid-sized fully automated immunoassay instrument (Photo courtesy of Fujirebio Diagnostics).
Image: The LUMIPULSE G1200 is a robust mid-sized fully automated immunoassay instrument (Photo courtesy of Fujirebio Diagnostics).
The exact definition of sepsis has changed considerably over time, and remains controversial. This is probably due to the fact that sepsis develops as a continuum of severity, ranging from local infection, through bloodstream propagation to septic shock.

Due to the steadily increasing use of procalcitonin (PCT) measurement for diagnosis and management of bacterial infections, the number of PCT immunoassays available on the diagnostic market has increased in recent years, with automated immunoassays replacing the former manual techniques.

Clinical biochemists at the University of Verona carried out an analytical evaluation of a two-step sandwich immunoassay for PCT in human serum and plasma. This analytical evaluation encompassed the calculation of the limit of blank (LOB), limit of detection (LOD), functional sensitivity, intra- and inter-assay imprecision, confirmation of linearity and a comparison with the ELFA immunoassay Vidas BRAHMS PCT.

The assay evaluated was the Lumipulse G BRAHMS PCT immunoassay adapted to be used on the LUMIPULSE G1200 system. The assay is based on ChemiLuminescent Enzyme Immunoassay (CLEIA) technology. Briefly, the PCT molecules present in the test sample bind to monoclonal anti-PCT mouse antibodies and to capture anti-calcitonin mouse antibodies coated on polystyrene beads, thus generating stable immune complexes and the luminescence is monitored at 477 nm. The assay displays an analytical sensitivity of 0.0048 ng/mL, a functional sensitivity of 0.0079 ng/mL and linearity comprised between 0.02 to 85.1 ng/mL and results are available in 30 minutes.

The LOB, LOD and functional sensitivity were 0.001 ng/mL, 0.0016 ng/mL and 0.008 ng/mL, respectively. The total analytical imprecision was found to be 2.1% and the linearity was excellent in the range of concentrations between 0.006 to 75.5 ng/mL. The correlation coefficient with Vidas BRAHMS PCT was 0.995. The diagnostic agreement was 100% at 0.5 ng/mL, 97% at 2.0 ng/mL and 95% at 10 ng/mL, respectively.

The authors concluded that the results of the evaluation of Lumipulse G BRAHMS PCT demonstrate that this technique exhibits excellent analytical performance, among the best of the methods currently available on the diagnostic market for routine PCT measurement. However, the significant bias shown by comparing the two immunoassays studied emphasizes that longitudinal patient monitoring should be always performed using the same immunoassay. The study was published on December 1, 2016, in the journal Practical Laboratory Medicine.

Platinum Member
COVID-19 Rapid Test
OSOM COVID-19 Antigen Rapid Test
Magnetic Bead Separation Modules
MAG and HEATMAG
Complement 3 (C3) Test
GPP-100 C3 Kit
Gold Member
Real-time PCR System
GentierX3 Series

Print article

Channels

Clinical Chemistry

view channel
Image: The 3D printed miniature ionizer is a key component of a mass spectrometer (Photo courtesy of MIT)

3D Printed Point-Of-Care Mass Spectrometer Outperforms State-Of-The-Art Models

Mass spectrometry is a precise technique for identifying the chemical components of a sample and has significant potential for monitoring chronic illness health states, such as measuring hormone levels... Read more

Hematology

view channel
Image: The CAPILLARYS 3 DBS devices have received U.S. FDA 510(k) clearance (Photo courtesy of Sebia)

Next Generation Instrument Screens for Hemoglobin Disorders in Newborns

Hemoglobinopathies, the most widespread inherited conditions globally, affect about 7% of the population as carriers, with 2.7% of newborns being born with these conditions. The spectrum of clinical manifestations... Read more

Microbiology

view channel
Image: The ePlex system has been rebranded as the cobas eplex system (Photo courtesy of Roche)

Enhanced Rapid Syndromic Molecular Diagnostic Solution Detects Broad Range of Infectious Diseases

GenMark Diagnostics (Carlsbad, CA, USA), a member of the Roche Group (Basel, Switzerland), has rebranded its ePlex® system as the cobas eplex system. This rebranding under the globally renowned cobas name... Read more

Pathology

view channel
Image: The revolutionary autonomous blood draw technology is witnessing growing demands (Photo courtesy of Vitestro)

Robotic Blood Drawing Device to Revolutionize Sample Collection for Diagnostic Testing

Blood drawing is performed billions of times each year worldwide, playing a critical role in diagnostic procedures. Despite its importance, clinical laboratories are dealing with significant staff shortages,... Read more
Copyright © 2000-2024 Globetech Media. All rights reserved.