Malaria Test Assessed in Nonendemic Setting

By LabMedica International staff writers
Posted on 23 May 2011
A rapid diagnostic test (RDT) for malaria has been evaluated in a standard reference laboratory using stored and fresh blood samples. More...


In nonendemic settings, where microscopic expertise is lacking due to low incidence, malaria RDTs are of value for the diagnosis of malaria and they provide information about the possibly fatal Plasmodium falciparum infections.

Scientists at Maastricht University, (Maastricht, The Netherlands), performed both a retrospective study on 341 stored blood samples and prospective study. In the prospective study, the SD Bioline Malaria Ag 05FK40, (SDFK40), RDT was used on 181 fresh samples and run side by side with two other RDTs, OptiMAL and SDFK60. The SDFK40 is a lateral flow antigen-detection test in a cassette format, targeting P. falciparum-specific parasite lactate dehydrogenase (Pf-pLDH) and pan-Plasmodium LDH, (pan-pLDH). The prospective part was performed between March 2009 and October 2010, and included first samples of each patient for which microscopy or one of the both routinely used RDTs (OptiMAL and SDFK60) were positive for malaria.

Overall sensitivities for P. falciparum tested retrospectively and prospectively were 67.9% and 78.8%, reaching 100% and 94.6% at parasite densities greater than 1,000/μL. Sensitivity at parasite densities less than 100/μL was 9.1%. Overall sensitivities for P. vivax and P. ovale were 86.7% and 80.0% (retrospectively) and 92.9% and 76.9% (prospectively), reaching 94.7% for both species (retrospective panel) at parasite densities greater than 500/μL. Sensitivity for P. malariae was 21.4%. None of the Plasmodium negative samples in the retrospective panel reacted positive.

The authors concluded that SDFK40 RDT (Standard Diagnostics Inc.; Suwon, Korea), performed exceptionally for P. falciparum samples at high parasite densities as well as for detection of P. vivax and P. ovale at parasite densities greater than 500/μL. The SDFK40 was evaluated as easy to use, with an excellent clearance of the background and clearly visible test lines. Compared to OptiMAL Rapid Malaria test, (Diamed AG; Cressier, Switzerland), and Standard Diagnostics other RDT, called SD Bioline Ag Pf/Pan 05FK60 (SDFK60), SDFK40 showed lower sensitivities for P. falciparum, but better detection of P. ovale. The study was published online on January 12, 2011, in the Malaria Journal.

Related Links:
Maastricht University
Standard Diagnostics Inc.
Diamed AG



Platinum Member
Automated Coagulation Analyzer
Hemolumi H6
Gold Member
Aspiration System
VACUSAFE
Electrolyte Analyzer
BKE-B
New
Fully Automated Urinalysis System
DxU 1800 Fully Automated Urinalysis System
Read the full article by registering today, it's FREE! It's Free!
Register now for FREE to LabMedica.com and get access to news and events that shape the world of Clinical Laboratory Medicine.
  • Free digital version edition of LabMedica International sent by email on regular basis
  • Free print version of LabMedica International magazine (available only outside USA and Canada).
  • Free and unlimited access to back issues of LabMedica International in digital format
  • Free LabMedica International Newsletter sent every week containing the latest news
  • Free breaking news sent via email
  • Free access to Events Calendar
  • Free access to LinkXpress new product services
  • REGISTRATION IS FREE AND EASY!
Click here to Register








Channels

Clinical Chemistry

view channel
Image Credit: 123RF

Screening After Pneumococcal Disease May Reveal Undiagnosed Blood Cancer or Immune Disorders

Severe pneumococcal disease requiring hospitalization often presents as pneumonia, particularly in older adults and people with cancer or compromised immune function. Because M protein testing and antibody... Read more

Technology

view channel
Image: The laser-based photoacoustic spectroscopy setup consists of a Mid-IR laser equipped with three QCL modules covering wavelengths from 5.6 μm to 12.9 μm, two silver coated mirrors (SCM), a dichroic mirror (DM) with a transmittance of 90%, a thermal power sensor head (PM) to monitor the output laser power, a mechanical chopper (MC) for frequency modulation and a CEPAS-detector with a self-designed swab holder (SH). (Credit: Graunke, T., Scholz, T., Pieniak, M. et al. Scientific Reports (2026). https://doi.org/10.1038/s41598-026-68298-9)

Laser-Based Swab Analysis Shows Promise for Detecting Disease-Linked Odor Patterns

Disease-related changes in volatile organic compounds can alter body odor, producing measurable patterns in exhaled breath and bodily fluids. Current analytical methods can be complex, time-consuming,... Read more

Industry

view channel
Image

Collaboration Combines AI Cognitive Assessment and RNA Blood Testing for Earlier Alzheimer’s Detection

Alzheimer’s disease is often identified only after substantial neurodegeneration, partly because current diagnostic pathways are fragmented and difficult to scale. As treatment shifts toward earlier intervention,... Read more
Copyright © 2000-2026 Globetech Media. All rights reserved.