Plasma Lactate Measurement Used to Test Variability

By LabMedica International staff writers
Posted on 05 Nov 2018
Print article
Image: The Cobas 6000 analyzer series for clinical chemistry and immunochemistry assays (Photo courtesy of Roche Diagnostics).
Image: The Cobas 6000 analyzer series for clinical chemistry and immunochemistry assays (Photo courtesy of Roche Diagnostics).
Clinical laboratories are required to adhere to manufacturers’ recommended sample handling instructions as part of their compliance with accreditation certification. However, manufacturer's recommendations do not always reflect current practice and deviations from recommendations necessitate extensive validation studies.

Lactate is an intermediary metabolite produced in relatively anaerobic conditions. Measurement of circulating lactate level is useful in assessing tissue perfusion in critically ill patients and in those being suspected of sepsis. It is produced by many cells and circulating levels reflect rate of production and of metabolism by the body.

Laboratory scientists at Parkland Health and Hospital System (Dallas, TX, USA) and their colleagues collected triple blood samples from a total of 123 samples from 51 patients (23 females, 28 males) with age ranging from 23 to 83 (mean 53) years, being investigated for sepsis, hemodynamic and associated metabolic abnormalities and from 50 normal volunteers (37 females, 13 males) with age ranging from 31 to 59 (mean 50) years. One tube was transported on ice (4 °C); the others were maintained at room temperature (23 °C). Tubes stored at 4 °C were processed at 30 minutes from collection. Tubes stored at 23 °C were processed at 15 and at 30 minutes from collection. Lactate levels were measured using Roche Diagnostics Cobas 6000 analyzer.

The team reported that lactate levels in normal subjects ranged from 0.6 to 3.1 mmol/L (median 1.1). Patient lactate levels ranged from 0.8 to 26.3 mmol/L (median 2.2). Bias in lactate levels following extended storage of samples from both normal subjects and patients ranged from −1.3 to 2.2 and from −1.0 to 1.0 mmol/L when stored for 30 minutes at 23 °C or at 4 °C, respectively. The bias between lactate levels at 30 minutes at 23 °C and 4 °C was −1.2 to −0.5 mmol/L for both populations. Although the bias was not statistically significant for all variables, a clinically significant (>0.2 mmol/L) bias was observed in 28% of normal and 7.0% of patient samples.

The authors concluded that achieving full compliance with recommended 15 minutes sample collection and handling time is unlikely to be met by most institutions and will thus impact lactate results. It would be helpful if manufacturers would consider, when developing guidelines on sample stability and on assays performance, constraints often seen in routine clinical practice. The study was published on October 24, 2018, in the journal Practical Laboratory Medicine.

Related Links:
Parkland Health and Hospital System

Platinum Member
COVID-19 Rapid Test
OSOM COVID-19 Antigen Rapid Test
Magnetic Bead Separation Modules
MAG and HEATMAG
Anti-Cyclic Citrullinated Peptide Test
GPP-100 Anti-CCP Kit
Gold Member
ADAMTS-13 Protease Activity Test
ATS-13 Activity Assay

Print article

Channels

Molecular Diagnostics

view channel
Image: Signs of multiple sclerosis show up in blood years before symptoms appear (Photo courtesy of vitstudio/Shutterstock)

Unique Autoantibody Signature to Help Diagnose Multiple Sclerosis Years before Symptom Onset

Autoimmune diseases such as multiple sclerosis (MS) are thought to occur partly due to unusual immune responses to common infections. Early MS symptoms, including dizziness, spasms, and fatigue, often... Read more

Hematology

view channel
Image: The CAPILLARYS 3 DBS devices have received U.S. FDA 510(k) clearance (Photo courtesy of Sebia)

Next Generation Instrument Screens for Hemoglobin Disorders in Newborns

Hemoglobinopathies, the most widespread inherited conditions globally, affect about 7% of the population as carriers, with 2.7% of newborns being born with these conditions. The spectrum of clinical manifestations... Read more

Immunology

view channel
Image: Exosomes can be a promising biomarker for cellular rejection after organ transplant (Photo courtesy of Nicolas Primola/Shutterstock)

Diagnostic Blood Test for Cellular Rejection after Organ Transplant Could Replace Surgical Biopsies

Transplanted organs constantly face the risk of being rejected by the recipient's immune system which differentiates self from non-self using T cells and B cells. T cells are commonly associated with acute... Read more

Microbiology

view channel
Image: Microscope image showing human colorectal cancer tumor with Fusobacterium nucleatum stained in a red-purple color (Photo courtesy of Fred Hutch Cancer Center)

Mouth Bacteria Test Could Predict Colon Cancer Progression

Colon cancer, a relatively common but challenging disease to diagnose, requires confirmation through a colonoscopy or surgery. Recently, there has been a worrying increase in colon cancer rates among younger... Read more

Pathology

view channel
Image: A new study has identified patterns that predict ovarian cancer relapse (Photo courtesy of Cedars-Sinai)

Spatial Tissue Analysis Identifies Patterns Associated With Ovarian Cancer Relapse

High-grade serous ovarian carcinoma is the most lethal type of ovarian cancer, and it poses significant detection challenges. Typically, patients initially respond to surgery and chemotherapy, but the... Read more
Copyright © 2000-2024 Globetech Media. All rights reserved.