Plasma Lactate Measurement Used to Test Variability

By LabMedica International staff writers
Posted on 05 Nov 2018
Print article
Image: The Cobas 6000 analyzer series for clinical chemistry and immunochemistry assays (Photo courtesy of Roche Diagnostics).
Image: The Cobas 6000 analyzer series for clinical chemistry and immunochemistry assays (Photo courtesy of Roche Diagnostics).
Clinical laboratories are required to adhere to manufacturers’ recommended sample handling instructions as part of their compliance with accreditation certification. However, manufacturer's recommendations do not always reflect current practice and deviations from recommendations necessitate extensive validation studies.

Lactate is an intermediary metabolite produced in relatively anaerobic conditions. Measurement of circulating lactate level is useful in assessing tissue perfusion in critically ill patients and in those being suspected of sepsis. It is produced by many cells and circulating levels reflect rate of production and of metabolism by the body.

Laboratory scientists at Parkland Health and Hospital System (Dallas, TX, USA) and their colleagues collected triple blood samples from a total of 123 samples from 51 patients (23 females, 28 males) with age ranging from 23 to 83 (mean 53) years, being investigated for sepsis, hemodynamic and associated metabolic abnormalities and from 50 normal volunteers (37 females, 13 males) with age ranging from 31 to 59 (mean 50) years. One tube was transported on ice (4 °C); the others were maintained at room temperature (23 °C). Tubes stored at 4 °C were processed at 30 minutes from collection. Tubes stored at 23 °C were processed at 15 and at 30 minutes from collection. Lactate levels were measured using Roche Diagnostics Cobas 6000 analyzer.

The team reported that lactate levels in normal subjects ranged from 0.6 to 3.1 mmol/L (median 1.1). Patient lactate levels ranged from 0.8 to 26.3 mmol/L (median 2.2). Bias in lactate levels following extended storage of samples from both normal subjects and patients ranged from −1.3 to 2.2 and from −1.0 to 1.0 mmol/L when stored for 30 minutes at 23 °C or at 4 °C, respectively. The bias between lactate levels at 30 minutes at 23 °C and 4 °C was −1.2 to −0.5 mmol/L for both populations. Although the bias was not statistically significant for all variables, a clinically significant (>0.2 mmol/L) bias was observed in 28% of normal and 7.0% of patient samples.

The authors concluded that achieving full compliance with recommended 15 minutes sample collection and handling time is unlikely to be met by most institutions and will thus impact lactate results. It would be helpful if manufacturers would consider, when developing guidelines on sample stability and on assays performance, constraints often seen in routine clinical practice. The study was published on October 24, 2018, in the journal Practical Laboratory Medicine.

Related Links:
Parkland Health and Hospital System

Platinum Member
COVID-19 Rapid Test
OSOM COVID-19 Antigen Rapid Test
One Step HbA1c Measuring System
GREENCARE A1c
Complement 3 (C3) Test
GPP-100 C3 Kit
New
Gold Member
Fully Automated Cell Density/Viability Analyzer
BioProfile FAST CDV

Print article

Channels

Molecular Diagnostics

view channel
Image: The study showed the blood-based cancer screening test detects 83% of people with colorectal cancer with specificity of 90% (Photo courtesy of Guardant Health)

Blood Test Shows 83% Accuracy for Detecting Colorectal Cancer

Colorectal cancer is the second biggest cause of cancer deaths among adults in the U.S., with forecasts suggesting 53,010 people might die from it in 2024. While fewer older adults are dying from this... Read more

Hematology

view channel
Image: The Gazelle Hb Variant Test (Photo courtesy of Hemex Health)

First Affordable and Rapid Test for Beta Thalassemia Demonstrates 99% Diagnostic Accuracy

Hemoglobin disorders rank as some of the most prevalent monogenic diseases globally. Among various hemoglobin disorders, beta thalassemia, a hereditary blood disorder, affects about 1.5% of the world's... Read more

Microbiology

view channel
Image: The new platform is designed to perform blood-based diagnoses of nontuberculosis mycobacteria (Photo courtesy of 123RF)

New Blood Test Cuts Diagnosis Time for Nontuberculous Mycobacteria Infections from Months to Hours

Breathing in nontuberculous mycobacteria (NTM) is a common experience for many people. These bacteria are present in water systems, soil, and dust all over the world and usually don't cause any problems.... Read more

Industry

view channel
Image: These new assays are being developed for use on the recently introduced DxI 9000 Immunoassay Analyzer (Photo courtesy of Beckman Coulter)

Beckman Coulter and Fujirebio Expand Partnership on Neurodegenerative Disease Diagnostics

Beckman Coulter Diagnostics (Brea, CA, USA) and Fujirebio Diagnostics (Tokyo, Japan) have expanded their partnership focused on the development, manufacturing and clinical adoption of neurodegenerative... Read more
Copyright © 2000-2024 Globetech Media. All rights reserved.