Liquid Chromatography Analyzer Evaluated for Hemoglobinopathy

By LabMedica International staff writers
Posted on 18 Jan 2017
Print article
Image: The ADAMS A1c, HA-8180 is a fully automated Glycohemoglobin (HbA1c) analyzer based on HPLC (Photo courtesy of Arkray).
Image: The ADAMS A1c, HA-8180 is a fully automated Glycohemoglobin (HbA1c) analyzer based on HPLC (Photo courtesy of Arkray).
The quantification of HbA2 and HbF, and the detection and quantification of hemoglobin variants, is an essential tool in the diagnosis of hemoglobinopathies such as thalassemia or sickle cell syndromes.

The narrow separation between normal and pathological HbA2 values, strict analytical quality of HbA2 measurement is an essential requirement for accurate diagnosis, particularly for genetic counseling when couples at-risk must be identified.

A hematologist at the Hospital Galdakao – Usansolo evaluated the analytical performance and quality of results obtained from a fully automated high-pressure liquid chromatography (HPLC) analyzer for routine estimation of HbA2 and the screening of β thalassemia. Blood specimens were obtained from patients whose diabetic control or hemoglobinopathy were being assessed.

The scientist measured HbA2 values with the fully automated Glycohemoglobin analyzer based on HPLC ADAMS A1c HA-8180 on Thalassemia mode and were compared with those obtained with their current method, the established HPLC ADAMSTM A1c HA-8160. Comparison was performed by running 400 samples from healthy subjects, 30 alpha and 80 beta carriers. Hemoglobin was measured on a XN counter. Different common variant hemoglobins were analyzed: HbS, HbC, HbD, HbJ, Hb Lepore, HbSC, Hb E, Hb O, and δβ thalassemia carriers with raised HbF. Twenty samples of heterozygous HbS were measured with capillary electrophoresis in the reference laboratory to discard the spurious increment of HbA2.

The mean difference between HbA2 results of ADAMS A1C HA-8180T and capillary electrophoresis, in samples of heterozygous HbS, was 0.2 %. The hematologist found an ample gap between healthy persons and carriers. A potential bias is that they have not included carriers of mutations producing borderline values of HbA2, which was absent in the area, and this fact must be considered in regions with high prevalence of those patients. The author concluded that the ADAMS A1c HA-8180T provided a rapid and reliable separation of HbA2. The measurement is accurate and reproducible, which is needed because of the slight difference between normal and pathological values. The gap in HbA2 values between normal subjects and β-thalassemia carriers makes this an appropriate method for rapid screening for carriers. The study was published in the December 2016 issue of the International Journal of Laboratory Hematology.

Platinum Member
COVID-19 Rapid Test
OSOM COVID-19 Antigen Rapid Test
Magnetic Bead Separation Modules
MAG and HEATMAG
Complement 3 (C3) Test
GPP-100 C3 Kit
New
Gold Member
TORCH Panel Rapid Test
Rapid TORCH Panel Test

Print article

Channels

Clinical Chemistry

view channel
Image: The 3D printed miniature ionizer is a key component of a mass spectrometer (Photo courtesy of MIT)

3D Printed Point-Of-Care Mass Spectrometer Outperforms State-Of-The-Art Models

Mass spectrometry is a precise technique for identifying the chemical components of a sample and has significant potential for monitoring chronic illness health states, such as measuring hormone levels... Read more

Molecular Diagnostics

view channel
Image: A network of inflammatory molecules may act as biomarker for risk of future cerebrovascular disease (Photo courtesy of 123RF)

Simple Blood Test Could Enable First Quantitative Assessments for Future Cerebrovascular Disease

Cerebral small vessel disease is a common cause of stroke and cognitive decline, particularly in the elderly. Presently, assessing the risk for cerebral vascular diseases involves using a mix of diagnostic... Read more

Immunology

view channel
Image: Exosomes can be a promising biomarker for cellular rejection after organ transplant (Photo courtesy of Nicolas Primola/Shutterstock)

Diagnostic Blood Test for Cellular Rejection after Organ Transplant Could Replace Surgical Biopsies

Transplanted organs constantly face the risk of being rejected by the recipient's immune system which differentiates self from non-self using T cells and B cells. T cells are commonly associated with acute... Read more

Microbiology

view channel
Image: The ePlex system has been rebranded as the cobas eplex system (Photo courtesy of Roche)

Enhanced Rapid Syndromic Molecular Diagnostic Solution Detects Broad Range of Infectious Diseases

GenMark Diagnostics (Carlsbad, CA, USA), a member of the Roche Group (Basel, Switzerland), has rebranded its ePlex® system as the cobas eplex system. This rebranding under the globally renowned cobas name... Read more

Pathology

view channel
Image: The Aperio GT 450 DX has received US FDA 510(k) clearance (Photo courtesy of Leica Biosystems)

Use of DICOM Images for Pathology Diagnostics Marks Significant Step towards Standardization

Digital pathology is rapidly becoming a key aspect of modern healthcare, transforming the practice of pathology as laboratories worldwide adopt this advanced technology. Digital pathology systems allow... Read more
Copyright © 2000-2024 Globetech Media. All rights reserved.