Automated Hemostasis Tests Affected By Hemolysis, Icterus And Lipemia

By LabMedica International staff writers
Posted on 02 Aug 2016
Print article
Image: The STA-Compact-Max viscosity-based detection coagulation analyzer (Photo courtesy of Stago).
Image: The STA-Compact-Max viscosity-based detection coagulation analyzer (Photo courtesy of Stago).
Laboratory diagnosis is more and more prominent in modern medicine and it is commonly accepted that approximately 70% of all medical decisions are based on the laboratory results and accurate results are therefore key for appropriate diagnosis.

One of the requirements for a clinical laboratory is that common interferences related to sample integrity such as hemolysis, icterus and lipemia (HIL) be evaluated with each reagent system and because of limited resources and budgetary constraints, the clinical laboratory relies on the manufacturer to document HIL estimates.

Scientists at the Royal Hallamshire Hospital (Sheffield, UK) and their French colleagues collected blood samples for testing plasma-based coagulation assays and molecular hemostasis assays. They assessed the influence of HIL on prothrombin time (PT), activated partial thromboplastin time (APTT) and fibrinogen assay using a viscosity-based detection analyzer (VBDS). Interference of hemolysis was studied in two different ways: spontaneous in vitro hemolysis judged to have occurred during sample collection transport or processing and spurious hemolysis.

The level of hemolysis was semi-quantitatively estimated based on the measurement of hemoglobin concentration using a XN-10 (Sysmex Corporation, Kobe, Japan). All the coagulation assays were performed using reagents and a STA-Compact-Max analyzer (Stago, Asnières sur Seine, France). One reagent, Stago’s STA-Liquid Fib, titrated human calcium thrombin, was used for the measurement of fibrinogen.

The scientists found that spontaneous hemolysis that occurred during sample collection and processing had no effect on PT with either a rabbit tissue factor extract or recombinant human tissue factor reagents. In contrast, addition of mechanically hemolysis cells impacted on PT for the highest hemoglobin concentration. For APTTs there was no significant difference between results in hemolysed and nonhemolysed samples. For the other two reagents studied, APTTs were statistically significantly shorter in hemolysed samples compared with nonhemolysed samples. This bias was clinically significant only for STA-PTT Automate reagent. For all three APTT reagents, the impact of hemolysis was sufficient to impact patient management decisions, and in some samples, the effects of lipemia and icterus were not clinically significant.

The authors concluded that their results confirm that PT and fibrinogen are not clinically significantly affected by HIL. The APTTs of some haemolysed samples were falsely normal with one reagent and more affected than two others. Hemolysed samples should be continuously rejected. Conversely, from a clinical standpoint, lipemia and icterus did not significantly affect APTT measured with the different reagents tested in combination with a VBDS analyzer.

Related Links:
Royal Hallamshire Hospital
Sysmex
Stago

Platinum Member
COVID-19 Rapid Test
OSOM COVID-19 Antigen Rapid Test
Magnetic Bead Separation Modules
MAG and HEATMAG
Anti-Cyclic Citrullinated Peptide Test
GPP-100 Anti-CCP Kit
Gold Member
Fully Automated Cell Density/Viability Analyzer
BioProfile FAST CDV

Print article

Channels

Clinical Chemistry

view channel
Image: The 3D printed miniature ionizer is a key component of a mass spectrometer (Photo courtesy of MIT)

3D Printed Point-Of-Care Mass Spectrometer Outperforms State-Of-The-Art Models

Mass spectrometry is a precise technique for identifying the chemical components of a sample and has significant potential for monitoring chronic illness health states, such as measuring hormone levels... Read more

Molecular Diagnostics

view channel
Image: A blood test could predict lung cancer risk more accurately and reduce the number of required scans (Photo courtesy of 123RF)

Blood Test Accurately Predicts Lung Cancer Risk and Reduces Need for Scans

Lung cancer is extremely hard to detect early due to the limitations of current screening technologies, which are costly, sometimes inaccurate, and less commonly endorsed by healthcare professionals compared... Read more

Immunology

view channel
Image: Exosomes can be a promising biomarker for cellular rejection after organ transplant (Photo courtesy of Nicolas Primola/Shutterstock)

Diagnostic Blood Test for Cellular Rejection after Organ Transplant Could Replace Surgical Biopsies

Transplanted organs constantly face the risk of being rejected by the recipient's immune system which differentiates self from non-self using T cells and B cells. T cells are commonly associated with acute... Read more

Pathology

view channel
Image: Comparison of traditional histopathology imaging vs. PARS raw data (Photo courtesy of University of Waterloo)

AI-Powered Digital Imaging System to Revolutionize Cancer Diagnosis

The process of biopsy is important for confirming the presence of cancer. In the conventional histopathology technique, tissue is excised, sliced, stained, mounted on slides, and examined under a microscope... Read more
Copyright © 2000-2024 Globetech Media. All rights reserved.