Coagulation Tests Affected by Sodium Citrate Concentration

By LabMedica International staff writers
Posted on 27 Aug 2018
Print article
Image: The STA COMPACT MAX 2 is a fully automated benchtop analyzer (Photo courtesy of Diagnostica Stago).
Image: The STA COMPACT MAX 2 is a fully automated benchtop analyzer (Photo courtesy of Diagnostica Stago).
Despite a high degree of standardization, coagulation tests largely depend on preanalytic factors. Apart from artifacts acquired during blood drawing and transportation, the in vitro anticoagulant used has a potential impact on the coagulation test result.

Usually, a 9:1 blood to sodium citrate ratio is used with a buffer range between 3.2% and 3.8%.The 3.2% buffered sodium citrate binds less assay-added calcium than 3.8% buffered sodium citrate, therefore clotting times tend to be shorter in 3.2% than 3.8% buffered sodium citrate. Although there are no exact data available, today 3.2% buffered citrate is widely used.

Medical Laboratory Scientists at the Medical University of Vienna (Vienna, Austria) carried out a prospective observational study with 76 volunteers, differences related to the citrate concentration were evaluated. For both buffer concentrations, reference range intervals were established. Using the same calibration settings, 3.2% and 3.8% buffered citrate samples were consecutively analyzed. The reagent-analyzer combinations were chosen with respect to measurement accuracy and validity, assays' robustness and practicability during the parameters' introduction phase at their laboratory.

The activated partial thromboplastin time lupus-sensitive reagent (aPTT-LA) for lupus anticoagulant (LAC) confirmation testing was analyzed by using plasma with and without a reagent containing phosphatidylethanolamine. Lupus anticoagulant confirmation testing was considered positive when the difference between both aPTT-LA clotting times was greater than eight seconds or the ratio between the neat diluted Russell viper venom time (dRVVT) and the dRVVTConfirm (dRVVTRatio, Staclot DRVV Screen/Staclot DRVV Confirm) was greater than 1.25.

The team found that most parameters evaluated presented good comparability between citrated samples taken with 3.2% and 3.8% trisodium buffer. The ellagic acid containing activated partial thromboplastin time reagent (aPTT-FS) indicated a systemic and proportional difference between both buffer concentrations, leading to an alteration in its reference ranges. Further, a confirmation test for lupus anticoagulant assessment (Staclot LA) showed only a moderate correlation with a proportional deviation between both citrate concentrations. Further, a statistically significant difference was found in the diluted Russell viper venom time confirmation testing, coagulation factors V and VIII, and the protein C activity, which was found to be of minor clinical relevance.

The authors concluded that the alteration of 3.2% and 3.8% buffered sodium citrate on specialized coagulation tests in their setting was of limited relevance. As an exception, the aPTT-FS and aPTT-LADifference were significantly affected by the citrate concentration used. These findings demonstrate the comparability of data assessed with these citrate concentration ranges. The study was published in the August 2018 issue of the journal Archives of Pathology & Laboratory Medicine.

Related Links:
Medical University of Vienna

New
Gold Member
Human Chorionic Gonadotropin Test
hCG Quantitative - R012
Verification Panels for Assay Development & QC
Seroconversion Panels
New
Herpes Simplex Virus ELISA
HSV 2 IgG – ELISA
New
Silver Member
Fixed Speed Tube Rocker
GTR-FS

Print article

Channels

Clinical Chemistry

view channel
Image: Professor Nicole Strittmatter (left) and first author Wei Chen stand in front of the mass spectrometer with a tissue sample (Photo courtesy of Robert Reich/TUM)

Mass Spectrometry Detects Bacteria Without Time-Consuming Isolation and Multiplication

Speed and accuracy are essential when diagnosing diseases. Traditionally, diagnosing bacterial infections involves the labor-intensive process of isolating pathogens and cultivating bacterial cultures,... Read more

Molecular Diagnostics

view channel
Image: Health Canada has approved SPINEstat, a first-in-class diagnostic blood test for axSpA, as a Class II medical device (Photo courtesy of Augurex)

First-in-Class Diagnostic Blood Test Detects Axial Spondyloarthritis

Axial spondyloarthritis (axSpA) is a chronic inflammatory autoimmune condition that typically affects individuals during their most productive years, with symptoms often emerging before the age of 45.... Read more

Immunology

view channel
Image: The cancer stem cell test can accurately choose more effective treatments (Photo courtesy of University of Cincinnati)

Stem Cell Test Predicts Treatment Outcome for Patients with Platinum-Resistant Ovarian Cancer

Epithelial ovarian cancer frequently responds to chemotherapy initially, but eventually, the tumor develops resistance to the therapy, leading to regrowth. This resistance is partially due to the activation... Read more

Technology

view channel
Image: The new algorithms can help predict which patients have undiagnosed cancer (Photo courtesy of Adobe Stock)

Advanced Predictive Algorithms Identify Patients Having Undiagnosed Cancer

Two newly developed advanced predictive algorithms leverage a person’s health conditions and basic blood test results to accurately predict the likelihood of having an undiagnosed cancer, including ch... Read more

Industry

view channel
Image: The collaboration aims to leverage Oxford Nanopore\'s sequencing platform and Cepheid\'s GeneXpert system to advance the field of sequencing for infectious diseases (Photo courtesy of Cepheid)

Cepheid and Oxford Nanopore Technologies Partner on Advancing Automated Sequencing-Based Solutions

Cepheid (Sunnyvale, CA, USA), a leading molecular diagnostics company, and Oxford Nanopore Technologies (Oxford, UK), the company behind a new generation of sequencing-based molecular analysis technologies,... Read more
Copyright © 2000-2025 Globetech Media. All rights reserved.