Serum Tubes Reduce False Positive Troponin Results

By LabMedica International staff writers
Posted on 04 Jul 2012
Blood samples collected in both Rapid Serum Tubes (RST) and Plasma Separator Tubes (PST) have been compared for cardiac troponin analysis. More...


Plasma samples are often used for cardiac troponin testing because they allow for faster turnaround time compared to serum separator tubes (SST), which require up to 30 minutes to clot.

Scientists at the Mayo Clinic (Rochester MN, USA) measured cardiac troponin (cTnT) concentrations in 443 emergency department patients who had a cardiac biomarker panel ordered and had both RST and PST specimens collected. All cTnT analysis was performed using the 4th generation Roche Troponin T STAT assay on a single Roche Cobas e411 analyzer (Roche Diagnostics, Indianapolis, IN, USA).

The serum RST and plasma PST collection tubes used are products of Becton–Dickinson and Company (Franklin Lakes, NJ, USA). The investigators determined the mean bias between the RST and PST cTnT concentrations, as well as the clinical concordance between RST and PST results. Clinically concordant results were defined as cTnT results either both being undetectable at less than 0.01 ng/mL, or if detectable, equal or greater than 0.01 ng/mL, a difference of equal or greater than 0.02 ng/mL between specimen types.

In 328 of 443 samples, cTnT concentrations were undetectable in both the RST and PST tubes. Seven samples had a detectable cTnT result from the PST tube, but had undetectable cTnT in the RST tube collected simultaneously. Rare false positive results in lithium heparin samples were not observed with rapid clot serum tubes. Clinically discordant cTnT results were observed between PST and RST in 10 of the 443 patient samples, with three of these being very elevated cTnT results to the degree that the diagnosis would have not been affected if cTnT results were used from either PST or RST samples.

The authors concluded that that samples collected in RST tubes yield cTnT results, which are analytically and clinically concordant to samples collected in PST tubes. Similar to studies of Troponin I on the Beckman-Coulter (Brea, CA, USA) platform they found that use of RST for collection of cardiac biomarker specimens may reduce the incidence of false positive Roche Troponin T results. The study was published in the July 2012 issue of the journal Clinical Biochemistry.

Related Links:

Roche Diagnostics
Becton–Dickinson and Company
Beckman–Coulter



Gold Member
Quantitative POC Immunoassay Analyzer
EASY READER+
Online QC Software
Acusera 24•7
Manual Pipetting Aid
Pipette Controllers macro
HPV Test
Allplex HPV28 Detection
Read the full article by registering today, it's FREE! It's Free!
Register now for FREE to LabMedica.com and get access to news and events that shape the world of Clinical Laboratory Medicine.
  • Free digital version edition of LabMedica International sent by email on regular basis
  • Free print version of LabMedica International magazine (available only outside USA and Canada).
  • Free and unlimited access to back issues of LabMedica International in digital format
  • Free LabMedica International Newsletter sent every week containing the latest news
  • Free breaking news sent via email
  • Free access to Events Calendar
  • Free access to LinkXpress new product services
  • REGISTRATION IS FREE AND EASY!
Click here to Register








Channels

Hematology

view channel
Image Credit: Shutterstock

New Biomarkers Predict Resistance to Targeted Therapy in Rare Blood Cancer

Blastic plasmacytoid dendritic cell neoplasm (BPDCN) is a rare and aggressive leukemia with limited treatment options and a poor prognosis. Although tagraxofusp is the first approved targeted therapy for... Read more

Immunology

view channel
Image:Proteomic tear-fluid analysis revealed abnormal patterns in proteins that regulate nerves and T cells in individuals with eye problems (Image Credit: Adobe Stock)

Diagnostic Models Detect Hidden Eye Abnormalities After Mild COVID-19

Persistent ocular symptoms after COVID-19 can severely affect reading, work, and daily tasks, yet standard eye exams often reveal no clear abnormalities. Patients experiencing photophobia, eye pain, and... Read more

Industry

view channel
Photo courtesy of Natera

Natera’s Signatera Earns IVDR Certification for Solid Tumor MRD Testing

Natera’s Signatera has received certification as a Class C device under the European Union’s In Vitro Diagnostic Regulation (IVDR), becoming the first personalized MRD test for solid tumors to achieve... Read more
Copyright © 2000-2026 Globetech Media. All rights reserved.