Serum Tubes Reduce False Positive Troponin Results

By LabMedica International staff writers
Posted on 04 Jul 2012
Blood samples collected in both Rapid Serum Tubes (RST) and Plasma Separator Tubes (PST) have been compared for cardiac troponin analysis. More...


Plasma samples are often used for cardiac troponin testing because they allow for faster turnaround time compared to serum separator tubes (SST), which require up to 30 minutes to clot.

Scientists at the Mayo Clinic (Rochester MN, USA) measured cardiac troponin (cTnT) concentrations in 443 emergency department patients who had a cardiac biomarker panel ordered and had both RST and PST specimens collected. All cTnT analysis was performed using the 4th generation Roche Troponin T STAT assay on a single Roche Cobas e411 analyzer (Roche Diagnostics, Indianapolis, IN, USA).

The serum RST and plasma PST collection tubes used are products of Becton–Dickinson and Company (Franklin Lakes, NJ, USA). The investigators determined the mean bias between the RST and PST cTnT concentrations, as well as the clinical concordance between RST and PST results. Clinically concordant results were defined as cTnT results either both being undetectable at less than 0.01 ng/mL, or if detectable, equal or greater than 0.01 ng/mL, a difference of equal or greater than 0.02 ng/mL between specimen types.

In 328 of 443 samples, cTnT concentrations were undetectable in both the RST and PST tubes. Seven samples had a detectable cTnT result from the PST tube, but had undetectable cTnT in the RST tube collected simultaneously. Rare false positive results in lithium heparin samples were not observed with rapid clot serum tubes. Clinically discordant cTnT results were observed between PST and RST in 10 of the 443 patient samples, with three of these being very elevated cTnT results to the degree that the diagnosis would have not been affected if cTnT results were used from either PST or RST samples.

The authors concluded that that samples collected in RST tubes yield cTnT results, which are analytically and clinically concordant to samples collected in PST tubes. Similar to studies of Troponin I on the Beckman-Coulter (Brea, CA, USA) platform they found that use of RST for collection of cardiac biomarker specimens may reduce the incidence of false positive Roche Troponin T results. The study was published in the July 2012 issue of the journal Clinical Biochemistry.

Related Links:

Roche Diagnostics
Becton–Dickinson and Company
Beckman–Coulter



Platinum Member
Automated Coagulation Analyzer
Hemolumi H6
βeta-Hydroxybutyrate Reagent
β-Hydroxybutyrate LiquiColor®
Manual Pipetting Aid
Pipette Controllers macro
Gold Member
NEW PRODUCT : SILICONE WASHING MACHINE TRAY COVER WITH VICOLAB SILICONE NET VICOLAB®
REGISTRED 682.9
Read the full article by registering today, it's FREE! It's Free!
Register now for FREE to LabMedica.com and get access to news and events that shape the world of Clinical Laboratory Medicine.
  • Free digital version edition of LabMedica International sent by email on regular basis
  • Free print version of LabMedica International magazine (available only outside USA and Canada).
  • Free and unlimited access to back issues of LabMedica International in digital format
  • Free LabMedica International Newsletter sent every week containing the latest news
  • Free breaking news sent via email
  • Free access to Events Calendar
  • Free access to LinkXpress new product services
  • REGISTRATION IS FREE AND EASY!
Click here to Register








Channels

Molecular Diagnostics

view channel
Image: A new study links broad genomic sequencing before first-line therapy to longer survival, supporting its role in treatment decisions (Image Credit: Shutterstock)

Study Supports Broad Genomic Profiling Before First-Line Lung Cancer Treatment

Biomarker testing plays a critical role in advanced non-small cell lung cancer by identifying patients who may benefit from targeted therapies. However, testing only one or two genes can miss actionable... Read more

Immunology

view channel
Image: The findings underscore the importance of aligning allergy testing with clinical history and diagnostic guidelines to reduce inappropriate testing and its consequences (Image Credit: iStock)

Survey Finds Gap Between Food Allergy Testing Awareness and Clinical Practice

Food allergy evaluation can be challenging when laboratory results do not match clinical symptoms. Broad food allergy panels may detect sensitization without confirming an actual allergy, potentially leading... Read more

Industry

view channel
Image Credit: iStock

C2N Makes PrecivityAD2 Alzheimer’s Blood Test Available to Clinicians

Diagnosing Alzheimer’s disease can be challenging, particularly in patients with mild cognitive impairment or early dementia. Assessing brain amyloid pathology can guide clinical decisions, but diagnostic... Read more
Copyright © 2000-2026 Globetech Media. All rights reserved.