Fragmented Red Blood Cell Measurement Automated

By LabMedica International staff writers
Posted on 02 Jul 2012
The laboratory diagnosis of thrombotic microangiopathies (TMA) or disorders that may mimic their features remains difficult. More...


The detection of schistocytes on the blood smear is a cornerstone finding to assess the diagnosis, but microscopic evaluation of schistocytes is still problematic with wide inter-observer variations.

Hematologists at the University Hospital (Nancy, France) conducted studies comparing automated and microscopic counts of Fragmented Red Blood Cells (FRC) and schistocytes. The data came from the day-to-day analysis of more than 90,000 peripheral blood (PB) samples for 10 years in four centers. More than 900 patients were examined because of a clinical suspicion of TMA and 17 samples were derived from preterm newborns who were supposed to give the highest FRC counts among nonpathologic samples.

FRC were counted by automated direct measurement using the two automated analyzers (ABCC): the ADVIA 120 analyzer (Siemens; Tarrytown, NY, USA) and the XE-2100 analyzer (Sysmex; Kobe, Japan). Schistocyte counting was carried out from May-Grünwald-Giemsa-stained PB smears. The reference range for FRC was less than 0.3% for the ADVIA and less than 0.5% for the Sysmex XE-2100.

The correlation coefficient for Siemens devices was around 0.7. The main problem resulted from an overestimation of FRC as compared to schistocytes from 0.4% to 0.5% for a count of schistocyte ranging from 0% to 1.5%. Conversely, the scientists observed an underestimation of FRC as compared to schistocytes subsequent to this 1.5% threshold. For the Sysmex XE-2100 counter, the overestimation of the FRC was more pronounced, at less 0.8%, and increased slightly with high schistocyte percentages. The presence of FRC below a threshold determined at 1% for both automated analyzers had a negative predictive value (PV) close to 100% to exclude the presence of schistocyte on the blood smear, but in relationship with a poor positive PV value.

The authors concluded that that systematic quantification of FRC might improve the laboratorians' ability to exclude samples with schistocytes. This should be especially useful among patients in whom TMA diagnosis is suspected. However, several limitations deserve consideration such as few false-negative FRC counts in a context of macrocytosis, great lack of specificity. Hence, a PB smear examination was still advised for the validation of the count if needed. The study was published on June 14, 2012, in the International Journal of Laboratory Hematology.

Related Links:

University Hospital
Siemens
Sysmex



Gold Member
Veterinary Hematology Analyzer
Exigo H400
Gold Member
Pharmacogenetics Panel
VeriDose Core Panel v2.0
New
Toxoplasma Gondii Test
Toxo IgG ELISA
New
cDNA Synthesis Kit
Ultimate cDNA Synthesis Kit
Read the full article by registering today, it's FREE! It's Free!
Register now for FREE to LabMedica.com and get access to news and events that shape the world of Clinical Laboratory Medicine.
  • Free digital version edition of LabMedica International sent by email on regular basis
  • Free print version of LabMedica International magazine (available only outside USA and Canada).
  • Free and unlimited access to back issues of LabMedica International in digital format
  • Free LabMedica International Newsletter sent every week containing the latest news
  • Free breaking news sent via email
  • Free access to Events Calendar
  • Free access to LinkXpress new product services
  • REGISTRATION IS FREE AND EASY!
Click here to Register








Channels

Molecular Diagnostics

view channel
Image: New biomarkers could someday make it easy to spot Parkinson’s disease in a patient’s blood sample (Photo courtesy of Shutterstock)

Unique Blood-Based Genetic Signature Can Diagnose Parkinson’s Disease

Parkinson's disease is primarily recognized for its impact on the central nervous system. Recent scientific progress has shifted focus to understanding the involvement of the immune system in the onset... Read more

Immunology

view channel
Image: Custom hardware and software for the real-time detection of immune cell biophysical signatures in NICU (Photo courtesy of Pediatric Research, DOI:10.1038/s41390-025-03952-y)

First-Of-Its-Kind Device Profiles Newborns' Immune Function Using Single Blood Drop

Premature infants are highly susceptible to severe and life-threatening conditions, such as sepsis and necrotizing enterocolitis (NEC). Newborn sepsis, which is a bloodstream infection occurring in the... Read more

Pathology

view channel
Image: The innovative doublet configuration and annular illumination overcome traditional metalens limitations (Photo courtesy of Tao Li and Jiacheng Sun/Nanjing University)

High-Resolution Metalens Doublet Microscope to Enhance Diagnostic Capabilities

Metalenses mark a groundbreaking leap in optical technology. Unlike traditional microscope objectives that rely on curved glass surfaces, metalenses utilize nanoscale structures to manipulate light at... Read more

Technology

view channel
Image: Concept of biosensor integrated into hygiene pads enabling direct semi-quantitative analysis of biomarkers in unprocessed menstruation blood (Photo courtesy of Dosnon, L et al. DOI: 10.1002/advs.202505170)

First Ever Technology Recognizes Disease Biomarkers Directly in Menstrual Blood in Sanitary Towels

Over 1.8 billion people menstruate worldwide, yet menstrual blood has been largely overlooked in medical practice. This blood contains hundreds of proteins, many of which correlate with their concentration... Read more
Copyright © 2000-2025 Globetech Media. All rights reserved.