Ion-Selective Electrode Methods Compared for Electrolytes

By LabMedica International staff writers
Posted on 02 Jul 2019
Electrolytes are among the most commonly requested clinical laboratory tests. More...
There are two methods for measuring electrolytes in plasma, both using the principle of an ion-selective electrode (ISE).

The first method, used in point-of-care blood gas analyzers, is direct potentiometry, which measures electrolyte activity in plasma without dilution. The second method, used in routine chemistry analyzers in central laboratories, is indirect ISE, which involves a pre-analytical serum or plasma dilution step.

Clinical biochemists at the Leiden University Medical Center (Leiden, The Netherlands) and their colleagues used anonymized laboratory results of patients admitted to the emergency room and intensive care unit, and demonstrated in four steps that this lack of interchangeability of sodium and chloride results can be attributed to differences in measurement principle, but also to analytical standardization issues of the blood gas analyzers.

The team extracted blood gas results performed on a RAPIDPoint 500 and routine chemistry performed on a Modular P800 sodium and chloride results for patients admitted to the emergency room of their hospital between June 1, 2014, and May 17, 2016.

The team investigated the effects of the introduction of the instrument factors in the blood gas analysers, by extracting blood gas and routine chemistry sodium and chloride results (with a maximum time of 1 hour between sample analysis) again for patients admitted to the emergency room of our hospital. This approach eliminated the systematic differences between blood gas and serum sodium for 92 emergency department patients, and reduced the mean difference in chloride to 1 mmol/L for 25 patients. The mean difference in sodium results between blood gas and serum for intensive care patients decreased from 4 mmol/L to 2 mmol/L.

The authors concluded that standardization of sodium and chloride across methods and matrices is not a simple task and emphasizes the need for a close collaboration between laboratory specialists and clinicians regarding education on pitfalls in laboratory medicine. The study was published on June 13, 2019, in the journal Practical Laboratory Medicine.

Related Links:
Leiden University Medical Center


Platinum Member
Automated Coagulation Analyzer
Hemolumi H6
βeta-Hydroxybutyrate Reagent
β-Hydroxybutyrate LiquiColor®
Manual Pipetting Aid
Pipette Controllers macro
Gold Member
Clinical Chemistry Assay
Sorbitol Dehydrogenase (SDH)
Read the full article by registering today, it's FREE! It's Free!
Register now for FREE to LabMedica.com and get access to news and events that shape the world of Clinical Laboratory Medicine.
  • Free digital version edition of LabMedica International sent by email on regular basis
  • Free print version of LabMedica International magazine (available only outside USA and Canada).
  • Free and unlimited access to back issues of LabMedica International in digital format
  • Free LabMedica International Newsletter sent every week containing the latest news
  • Free breaking news sent via email
  • Free access to Events Calendar
  • Free access to LinkXpress new product services
  • REGISTRATION IS FREE AND EASY!
Click here to Register








Channels

Molecular Diagnostics

view channel
Image: The test combines RNA signals from the tumor-immune microenvironment to predict treatment response, rather than relying on a single marker such as PD-L1 expression.(Image Credit: Adobe Stock)

RNA-Based Test Predicts Immunotherapy Response in Non-Small Cell Lung Cancer

Non-small cell lung cancer (NSCLC) represents the largest U.S. patient population eligible for immune checkpoint inhibitor therapy. Treatment selection often relies on PD-L1 expression, but a single protein... Read more

Immunology

view channel
Image Credit: Adobe Stock

Immune Profiling Identifies Early Maternal Autoantibodies Linked to Preterm Birth Risk

A typical pregnancy lasts about 40 weeks, but labor begins prematurely in roughly one in 10 pregnancies. Preterm birth is the leading cause of newborn death worldwide, yet its underlying causes and which... Read more

Technology

view channel
Image: The laser-based photoacoustic spectroscopy setup consists of a Mid-IR laser equipped with three QCL modules covering wavelengths from 5.6 μm to 12.9 μm, two silver coated mirrors (SCM), a dichroic mirror (DM) with a transmittance of 90%, a thermal power sensor head (PM) to monitor the output laser power, a mechanical chopper (MC) for frequency modulation and a CEPAS-detector with a self-designed swab holder (SH). (Credit: Graunke, T., Scholz, T., Pieniak, M. et al. Scientific Reports (2026). https://doi.org/10.1038/s41598-026-68298-9)

Laser-Based Swab Analysis Shows Promise for Detecting Disease-Linked Odor Patterns

Disease-related changes in volatile organic compounds can alter body odor, producing measurable patterns in exhaled breath and bodily fluids. Current analytical methods can be complex, time-consuming,... Read more

Industry

view channel
Image: NeoGenomics selected Indica Labs’ HALO AP platform to eable its digital pathology operations through a scalable, cloud-based framework (Photo courtesy of Indica Labs)

Expanded Partnership Scales AI-Powered Digital Pathology for Oncology

Oncology diagnostics rely on pathology and biomarker information to guide personalized treatment. As computational pathology and artificial intelligence (AI) expand insights from tissue analysis, laboratories... Read more
Copyright © 2000-2026 Globetech Media. All rights reserved.