Test Results of Fingerstick Blood Vary Significantly

By LabMedica International staff writers
Posted on 29 Nov 2015
Print article
Image: Blood obtained via fingerstick is commonly used in point-of-care assays (Photo courtesy of The Health).
Image: Blood obtained via fingerstick is commonly used in point-of-care assays (Photo courtesy of The Health).
Blood obtained via fingerstick is commonly used in point-of-care assays, but few studies have assessed variability in parameters obtained from successive drops of fingerstick blood, which may cause problems for clinical decision making and for assessing accuracy of point-of-care tests.

The most accurate way to carry out medial laboratory tests is to draw blood from a vein and send it to a laboratory but sometimes, such as in low-resource settings, results are needed more rapidly, or the health professional is not trained to draw vein-blood, or there is no laboratory. In such settings, there is a higher reliance on fingerstick tests.

Bioengineers at Rice University (Houston, TX, USA) used a hematology analyzer to analyze the hemoglobin concentration, total white blood cell (WBC) count, three-part WBC differential, and platelet count in six successive 20 µL of blood collected from one fingerstick from each of 11 donors. The team used a hemoglobinometer to measure the hemoglobin concentration of 10 drops of fingerstick blood from each of seven donors to check whether the minimum droplet size made a difference and they checked all the results against blood taken from donors' veins. They also followed best practice to ensure accurate results. For example, they wiped away the first droplet to remove traces of disinfectant, and they did not squeeze or "milk" the finger.

The results showed that hemoglobin content, platelet count and white blood cell count varied significantly from drop to drop. The average percent coefficient of variation (CV) for successive drops of fingerstick blood was higher by up to 3.4 times for hemoglobin, 5.7 times for WBC count, three times for lymphocyte count, 7.7 times for granulocyte count, and four times for platelets than in venous controls measured using a hematology analyzer. The average percent CV for fingerstick blood was up to five times higher for hemoglobin than venous blood measured using a point-of-care hemoglobinometer. The investigators found that averaging the results of six to nine successive droplet tests produced results on a par with the venous blood tests.

Meaghan M. Bond, a doctoral student and first author of the study, said, “In some donors, the hemoglobin concentration changed by more than 2 g/dL in the span of two successive drops of blood. Our results show that people need to take care to administer fingerstick tests in a way that produces accurate results because accuracy in these tests is increasingly important for diagnosing conditions like anemia, infections and sickle-cell anemia, malaria, HIV and other diseases.” The study was published on November 18, 2015, in the American Journal of Clinical Pathology.

Related Links:

Rice University


Platinum Member
COVID-19 Rapid Test
OSOM COVID-19 Antigen Rapid Test
Magnetic Bead Separation Modules
MAG and HEATMAG
Anti-Cyclic Citrullinated Peptide Test
GPP-100 Anti-CCP Kit
Gold Member
Systemic Autoimmune Testing Assay
BioPlex 2200 ANA Screen with MDSS

Print article

Channels

Clinical Chemistry

view channel
Image: The 3D printed miniature ionizer is a key component of a mass spectrometer (Photo courtesy of MIT)

3D Printed Point-Of-Care Mass Spectrometer Outperforms State-Of-The-Art Models

Mass spectrometry is a precise technique for identifying the chemical components of a sample and has significant potential for monitoring chronic illness health states, such as measuring hormone levels... Read more

Molecular Diagnostics

view channel
Image: A blood test could predict lung cancer risk more accurately and reduce the number of required scans (Photo courtesy of 123RF)

Blood Test Accurately Predicts Lung Cancer Risk and Reduces Need for Scans

Lung cancer is extremely hard to detect early due to the limitations of current screening technologies, which are costly, sometimes inaccurate, and less commonly endorsed by healthcare professionals compared... Read more

Immunology

view channel
Image: Exosomes can be a promising biomarker for cellular rejection after organ transplant (Photo courtesy of Nicolas Primola/Shutterstock)

Diagnostic Blood Test for Cellular Rejection after Organ Transplant Could Replace Surgical Biopsies

Transplanted organs constantly face the risk of being rejected by the recipient's immune system which differentiates self from non-self using T cells and B cells. T cells are commonly associated with acute... Read more

Microbiology

view channel
Image: The real-time multiplex PCR test is set to revolutionize early sepsis detection (Photo courtesy of Shutterstock)

1 Hour, Direct-From-Blood Multiplex PCR Test Identifies 95% of Sepsis-Causing Pathogens

Sepsis contributes to one in every three hospital deaths in the US, and globally, septic shock carries a mortality rate of 30-40%. Diagnosing sepsis early is challenging due to its non-specific symptoms... Read more

Pathology

view channel
Image: The QIAseq xHYB Mycobacterium tuberculosis Panel uses next-generation sequencing (Photo courtesy of 123RF)

New Mycobacterium Tuberculosis Panel to Support Real-Time Surveillance and Combat Antimicrobial Resistance

Tuberculosis (TB), the leading cause of death from an infectious disease globally, is a contagious bacterial infection that primarily spreads through the coughing of patients with active pulmonary TB.... Read more
Copyright © 2000-2024 Globetech Media. All rights reserved.