Newly Developed Point-of-Care Hemoglobinometer Evaluated

By LabMedica International staff writers
Posted on 10 Dec 2014
Print article
Image: The TrueHb hemometer testing kit (Photo courtesy of the Indian Institute of Technology Delhi).
Image: The TrueHb hemometer testing kit (Photo courtesy of the Indian Institute of Technology Delhi).
Hemoglobin (Hb) concentration is routinely measured using automated analyzers, and although these counters are very accurate and reliable, they are expensive and too cumbersome for portability.

Portable hemoglobinometer can potentially provide the solution to the challenges as they are affordable, precise, and convenient and require only a tiny sample of capillary or venous blood, do not require any calibration and display digital results immediately.

Scientists at the Indian Institute of Technology Delhi (IITD; New Delhi, India) collected 200 hundred randomly selected venous blood samples, submitted for hemogram study at the Department of Hematology, All India Institute of Medical Sciences (New Delhi, India). The hemoglobin values of these samples were evenly spread. There were 67, 76, and 57 samples in the ranges of 3.4–8 g/dL, 8–12.5 g/dL, and 12.5–20 g/dL respectively.

The investigators evaluated the performance of the portable hemometer called TrueHb (New Delhi, India) developed by IITD, and compared it with an automated five-part hematology analyzer intended for in vitro diagnostic use in clinical laboratories, the Sysmex counter XT 1800i (Sysmex; Kobe, Japan). The two set of values were comparatively analyzed and the repeatability of the performance of TrueHb was also evaluated against Sysmex value.

The scatter plot of TrueHb values and Sysmex values showed linear distribution with positive correlations. The intraclass correlation (ICC) values between the two set of values was found to be very high. The mean difference in Bland–Altman plots of TrueHb values against the Sysmex values was found to be −0.02, with limits of agreement between -0.777 and 0.732 g/dL. Statistical analysis suggested good repeatability in results of TrueHb, having a low mean coefficient of variation (CV) of 2.22, against 4.44, that of Sysmex values, and 95% confidence interval of 1.99 to 2.44, against 3.85 to 5.03, that of Sysmex values.

The authors concluded that a strong positive correlation between the two measurements devices suggest that the newly developed hemometer, TrueHb, can potentially replace a pathology laboratory analyzer for the purpose of measuring hemoglobin in blood samples. Keeping the performance of TrueHb in view, the newly developed device exhibited the potential to become a dependable tool for healthcare providers for measuring and monitoring hemoglobin levels of patients in conventional healthcare setups as well as in field studies. The study was published on November 22, 2014, in the International Journal of Laboratory Hematology.

Related Links:
Indian Institute of Technology Delhi 
All India Institute of Medical Sciences
TrueHb



Platinum Member
COVID-19 Rapid Test
OSOM COVID-19 Antigen Rapid Test
Magnetic Bead Separation Modules
MAG and HEATMAG
Complement 3 (C3) Test
GPP-100 C3 Kit
New
Gold Member
Plasma Control
Plasma Control Level 1

Print article

Channels

Clinical Chemistry

view channel
Image: The 3D printed miniature ionizer is a key component of a mass spectrometer (Photo courtesy of MIT)

3D Printed Point-Of-Care Mass Spectrometer Outperforms State-Of-The-Art Models

Mass spectrometry is a precise technique for identifying the chemical components of a sample and has significant potential for monitoring chronic illness health states, such as measuring hormone levels... Read more

Immunology

view channel
Image: A false color scanning election micrograph of lung cancer cells grown in culture (Photo courtesy of Anne Weston)

AI Tool Precisely Matches Cancer Drugs to Patients Using Information from Each Tumor Cell

Current strategies for matching cancer patients with specific treatments often depend on bulk sequencing of tumor DNA and RNA, which provides an average profile from all cells within a tumor sample.... Read more

Microbiology

view channel
Image: Microscope image showing human colorectal cancer tumor with Fusobacterium nucleatum stained in a red-purple color (Photo courtesy of Fred Hutch Cancer Center)

Mouth Bacteria Test Could Predict Colon Cancer Progression

Colon cancer, a relatively common but challenging disease to diagnose, requires confirmation through a colonoscopy or surgery. Recently, there has been a worrying increase in colon cancer rates among younger... Read more

Pathology

view channel
Image: Fingertip blood sample collection on the Babson Handwarmer (Photo courtesy of Babson Diagnostics)

Unique Hand-Warming Technology Supports High-Quality Fingertip Blood Sample Collection

Warming the hand is an effective way to facilitate blood collection from a fingertip, yet off-the-shelf solutions often do not fulfill laboratory requirements. Now, a unique hand-warming technology has... Read more
Copyright © 2000-2024 Globetech Media. All rights reserved.