Point-of-Care Hemoglobin Testing Compared with Laboratory Analyzer

By LabMedica International staff writers
Posted on 28 Jan 2014
Print article
Image: The noninvasive hemoglobin (SpHb) device (Photo image courtesy of Masimo Corp).
Image: The noninvasive hemoglobin (SpHb) device (Photo image courtesy of Masimo Corp).
Hemoglobin concentration is assessed to detect anemia and its associated morbidities and is usually determined from venous or capillary blood samples run on a laboratory analyzer.

The blood sample method requires a needle stick and results can be delayed, and exposes caregivers to risks associated with needle sticks and blood exposure. Noninvasive hemoglobin determination would be of benefit to patients and caregivers because it would allow for quick and painless point-of-care assessment.

Clinical scientists at Long Beach Veterans Administration Healthcare System (CA, USA) enrolled 474 subjects in a study to compare a noninvasive hemoglobin concentration measurement method with the traditional technique. A noninvasive hemoglobin (SpHb) test device was placed on the subject's nondominant ring finger. The sensor was covered with an opaque shield to prevent optical interference. A noninvasive hemoglobin measurement was obtained from each subject. Perfusion index, finger temperature, and oxygen saturation (SpO2), measured by the noninvasive hemoglobin test device, were also recorded.

Immediately following the noninvasive testing, a venous blood sample was obtained and analyzed for reference hemoglobin value (Hb) with a laboratory hematology analyzer (LH 500, Beckman Coulter; Brea, CA, USA). A capillary blood sample was also obtained by finger stick within 15 minutes of the noninvasive reading and the venous sample and analyzed immediately with a HemoCue 201 point-of-care hemoglobinometer (Cypress, CA, USA). The noninvasive hemoglobin (SpHb) test device was the Pronto-7, version 2.19, with a rainbow 4D sensor, rev E, (Masimo Corp., Irvine, CA, USA).

The reference hemoglobin concentration ranged from 8.6 g/dL to 17.4 g/dL. Analysis of data from all subjects showed a bias ± standard deviation of −0.1 ± 1.1 g/dL for Pronto-7 SpHb and −0.1 ± 1.6 g/dL for HemoCue, compared with the hematology analyzer. SpHb could be used to safety screen between 73% and 95% of the population to rule out anemia, with SpHb values of ≥11.5g/dL to 13 g/dL and an additional 0.5% of the population to rule in anemia, with SpHb values of 7.5g/dL to ≤10 g/dL.

The authors concluded that their study shows noninvasive hemoglobin measurement with SpHb offers similar accuracy as another established point-of-care method, HemoCue, in an outpatient clinic environment. Its noninvasive and quick testing characteristics may translate into added benefits to patient care. The study was published on January 13, 2014, in the International Journal of Laboratory Hematology.

Related Links:

Long Beach Veterans Administration Healthcare System
Beckman Coulter
Masimo Corp.


Platinum Member
COVID-19 Rapid Test
OSOM COVID-19 Antigen Rapid Test
Magnetic Bead Separation Modules
MAG and HEATMAG
Complement 3 (C3) Test
GPP-100 C3 Kit
Gold Member
Systemic Autoimmune Testing Assay
BioPlex 2200 ANA Screen with MDSS

Print article

Channels

Clinical Chemistry

view channel
Image: The 3D printed miniature ionizer is a key component of a mass spectrometer (Photo courtesy of MIT)

3D Printed Point-Of-Care Mass Spectrometer Outperforms State-Of-The-Art Models

Mass spectrometry is a precise technique for identifying the chemical components of a sample and has significant potential for monitoring chronic illness health states, such as measuring hormone levels... Read more

Molecular Diagnostics

view channel
Image: AI analysis of DNA fragmentomes and protein biomarkers noninvasively detects ovarian cancer (Photo courtesy of Adobe Stock)

Blood-Based Machine Learning Assay Noninvasively Detects Ovarian Cancer

Ovarian cancer is one of the most common causes of cancer deaths among women and has a five-year survival rate of around 50%. The disease is particularly lethal because it often doesn't cause symptoms... Read more

Immunology

view channel
Image: The groundbreaking treatment approach has shown promise in hard-to-treat cancers (Photo courtesy of 123RF)

Genetic Testing Combined With Personalized Drug Screening On Tumor Samples to Revolutionize Cancer Treatment

Cancer treatment typically adheres to a standard of care—established, statistically validated regimens that are effective for the majority of patients. However, the disease’s inherent variability means... Read more

Pathology

view channel
Image: The device can serve as a sample pretreatment tool for cytological diagnosis of malignant effusions (Photo courtesy of Microsystems & Nanoengineering: Zhu, Z., Ren, H., Wu, D. et al.)

Microfluidic Device for Cancer Detection Precisely Separates Tumor Entities

Tumor cell clusters are increasingly recognized as crucial in cancer pathophysiology, with growing evidence of their increased resistance to treatment and higher metastatic potential compared to single tumor cells.... Read more
Copyright © 2000-2024 Globetech Media. All rights reserved.