Routine Coagulation Testing Methods Compared

By Labmedica International staff writers
Posted on 13 Jun 2012


Coagulation testing can be estimated using two different detection methodologies; one is an optical method and the other a mechanical system.

Coagulation analyzers run the range of options, from automated instruments to manual devices, from high-volume processing to low-volume analysis, and from a wide selection of assays to a small, core group of tests.

Scientists at the Memorial Hospital (Istanbul, Turkey) assayed prothrombin time (PT) and activated partial thromboplastin time (aPTT) on the Coag-a-mate MTX II automated photooptical coagulation analyzer and the AMAX 200 automated photo-mechanical coagulation analyzer. A total of 424 patient samples were included in the study during a period of 15 days and plasma from 25 normal healthy individuals was assayed to obtain the geometric means and reference ranges for PT and aPTT for both the MTX II and AMAX 200 instruments which are marketed by TrinityBioTech (Berkeley Heights, NJ, USA).

The results of the coagulation tests showed good precision ranging between 0.7% and 1.8% coefficient of variation. Statistical analysis demonstrated an excellent correlation between the photooptical and mechanical analyzers for PT and aPTT. Plasma samples from 25 healthy normal donors yielded the following reference ranges: for PT, 10.5-12.3 seconds for MTX II and 11.3-13.4 seconds for the AMAX 200. The ranges for the aPTT were 24.5-33.9 seconds for MTX II and 25.5-35.8 seconds for the AMAX 200.

The authors concluded that all data obtained during the study indicate that the patient results obtained by the photooptical detection system are as reliable and statistically equivalent as those obtained using the mechanical detection system. Most notably, they also found that when testing suboptimal samples, such as those with hyperbilirubinemia or lipemia, the photooptical and mechanical detection methods are statistically equivalent. The study was published in the May 2012 issue of the Journal of Clinical Laboratory Analysis.

Related Links:

Memorial Hospital
TrinityBioTech




Flocked Swab Products
Flocked Swab Products
Glucose Analyzer
Glucose Analyzer
Electrolyte Analyzer
Electrolyte Analyzer

More Products

Latest Hematology News

LinkXpress
Click for LinkXpress
Reader Inquiry Service
Enter code to receive information:
Where I can find code?
Featured Whitepaper
SIEMENS DIAGNOSTICS :
Siemens Healthcare Diagnostics Anti-CCP: A Clinical Evaluation Whitepaper

Download Whitepaper
Events
ELA – European Lab Automation congress
06 Jun 2013 - 07 Jun 2013


European Human Genetics Conference 2013
08 Jun 2013 - 11 Jun 2013


8th Forum on Oxidative Stress and Aging
12 Jun 2013 - 14 Jun 2013


More events
GLOBETECH PUBLISHING
Latest Issue

View Digital Edition
Subscribe / Renew
DIASOURCE
77 ELEKTRONIKA
KARL HECHT GMBH & CO KG
BioPorto Diagnostics
GREINER-BIO-ONE
AACC
ALTO MARKETING LIMITED
  labmedica.com Copyright © 2000-2013 Globetech Media. All rights reserved.
Privacy Policy