Serum Tubes Reduce False Positive Troponin Results

By LabMedica International staff writers
Posted on 04 Jul 2012
Print article
Image: The Rapid Serum Tube (Photo courtesy of Becton–Dickinson and Company).
Image: The Rapid Serum Tube (Photo courtesy of Becton–Dickinson and Company).
Blood samples collected in both Rapid Serum Tubes (RST) and Plasma Separator Tubes (PST) have been compared for cardiac troponin analysis.

Plasma samples are often used for cardiac troponin testing because they allow for faster turnaround time compared to serum separator tubes (SST), which require up to 30 minutes to clot.

Scientists at the Mayo Clinic (Rochester MN, USA) measured cardiac troponin (cTnT) concentrations in 443 emergency department patients who had a cardiac biomarker panel ordered and had both RST and PST specimens collected. All cTnT analysis was performed using the 4th generation Roche Troponin T STAT assay on a single Roche Cobas e411 analyzer (Roche Diagnostics, Indianapolis, IN, USA).

The serum RST and plasma PST collection tubes used are products of Becton–Dickinson and Company (Franklin Lakes, NJ, USA). The investigators determined the mean bias between the RST and PST cTnT concentrations, as well as the clinical concordance between RST and PST results. Clinically concordant results were defined as cTnT results either both being undetectable at less than 0.01 ng/mL, or if detectable, equal or greater than 0.01 ng/mL, a difference of equal or greater than 0.02 ng/mL between specimen types.

In 328 of 443 samples, cTnT concentrations were undetectable in both the RST and PST tubes. Seven samples had a detectable cTnT result from the PST tube, but had undetectable cTnT in the RST tube collected simultaneously. Rare false positive results in lithium heparin samples were not observed with rapid clot serum tubes. Clinically discordant cTnT results were observed between PST and RST in 10 of the 443 patient samples, with three of these being very elevated cTnT results to the degree that the diagnosis would have not been affected if cTnT results were used from either PST or RST samples.

The authors concluded that that samples collected in RST tubes yield cTnT results, which are analytically and clinically concordant to samples collected in PST tubes. Similar to studies of Troponin I on the Beckman-Coulter (Brea, CA, USA) platform they found that use of RST for collection of cardiac biomarker specimens may reduce the incidence of false positive Roche Troponin T results. The study was published in the July 2012 issue of the journal Clinical Biochemistry.

Related Links:

Roche Diagnostics
Becton–Dickinson and Company
Beckman–Coulter


Platinum Member
COVID-19 Rapid Test
OSOM COVID-19 Antigen Rapid Test
Magnetic Bead Separation Modules
MAG and HEATMAG
Complement 3 (C3) Test
GPP-100 C3 Kit
Gold Member
Xylazine Immunoassay Test
Xylazine ELISA

Print article

Channels

Hematology

view channel
Image: The CAPILLARYS 3 DBS devices have received U.S. FDA 510(k) clearance (Photo courtesy of Sebia)

Next Generation Instrument Screens for Hemoglobin Disorders in Newborns

Hemoglobinopathies, the most widespread inherited conditions globally, affect about 7% of the population as carriers, with 2.7% of newborns being born with these conditions. The spectrum of clinical manifestations... Read more

Immunology

view channel
Image: A false color scanning election micrograph of lung cancer cells grown in culture (Photo courtesy of Anne Weston)

AI Tool Precisely Matches Cancer Drugs to Patients Using Information from Each Tumor Cell

Current strategies for matching cancer patients with specific treatments often depend on bulk sequencing of tumor DNA and RNA, which provides an average profile from all cells within a tumor sample.... Read more

Microbiology

view channel
Image: Microscope image showing human colorectal cancer tumor with Fusobacterium nucleatum stained in a red-purple color (Photo courtesy of Fred Hutch Cancer Center)

Mouth Bacteria Test Could Predict Colon Cancer Progression

Colon cancer, a relatively common but challenging disease to diagnose, requires confirmation through a colonoscopy or surgery. Recently, there has been a worrying increase in colon cancer rates among younger... Read more

Pathology

view channel
Image: Fingertip blood sample collection on the Babson Handwarmer (Photo courtesy of Babson Diagnostics)

Unique Hand-Warming Technology Supports High-Quality Fingertip Blood Sample Collection

Warming the hand is an effective way to facilitate blood collection from a fingertip, yet off-the-shelf solutions often do not fulfill laboratory requirements. Now, a unique hand-warming technology has... Read more
Copyright © 2000-2024 Globetech Media. All rights reserved.