Increased Mean Corpuscular Hemoglobin Concentration Scrutinized for Accuracy

By LabMedica International staff writers
Posted on 21 Dec 2016
Print article
Image: The XN-10 RET automated hematology system (Photo courtesy of Sysmex).
Image: The XN-10 RET automated hematology system (Photo courtesy of Sysmex).
In daily practice in hematology laboratories, spurious increased mean corpuscular hemoglobin concentration (MCHC) induces an analytical alarm and needs prompt corrective action to ensure delivery of the right results to the clinicians.

Elevated MCHC is a rare event in routine laboratory practice, but it must be managed properly. In daily practice, the MCHC limit defined by a specific commercial analyzer is fixed at 365 g/L. Exceeding this value leads to a suspicious ‘flag’ and this ‘flag’ has to be considered in an accreditation context to assess the accuracy of reported parameters.

Hematologists at the Hôpital de la Conception (Marseille, France) measured and analyzed in parallel with blood smears from 128 unknown patients with MCHC greater than 365 g/L, all erythrocyte parameters including reticulocyte parameters, chemistry index and osmolality. Differences between optical parameters (RBC-O, HGB-O) and usual parameters (RBC, HGB) obtained by impedance and photometry were also reported.

The scientists used the Sysmex XN-10 RET automated hematology system (Sysmex Corporation, Kobe Japan) that has two different technologies for achieving a full erythrocyte analysis. Erythrocytes are counted using an impedance method with a hydrodynamic focusing system in a fixed volume at room temperature. When required, XN-10 RET can provide a second erythrocyte count (RBC-O) using fluorescence flow cytometry after stabilization and warming at 41 °C in the incubation chamber. RBC-O is a measured parameter, corresponding to total erythrocyte count, including reticulocyte counts, whereas HGB-O is a calculated parameter derived mainly from the RBC-O count and RBC hemoglobin content (RBC-He).

The team classified four groups from their observations: 22 with red blood cell (RBC) agglutination; 17 with optical interference; 18 with RBC disease and 71 others including unclassified and/or patients with hyposmolar plasma. The use of RBC-O and HGB-O permitted efficient correction of the abnormalities when RBC agglutination and/or optical interference were present in 36 of 39 patients. Reticulocyte parameters permitted to elaborate an RBC score that allowed a highly sensitive detection of RBC disease patients (17/18).

The authors concluded that in case of elevated MCHC, their study proves the capability of XN-10 RET optical parameters to provide solutions in the majority of cases, especially concerning RBC cold agglutination and optical interference. The calculated RBC score offers a highly useful tool for managing a blood smear and specifying patients with RBC disease. This original study allows optimization of the workflow in laboratories eliminating manual tasks, guiding biological interpretation in the case of elevated MCHC. The study was originally published online on August 27, 2016, in the International Journal of Laboratory Hematology.

Related Links:
Hôpital de la Conception
Sysmex
Platinum Member
COVID-19 Rapid Test
OSOM COVID-19 Antigen Rapid Test
Magnetic Bead Separation Modules
MAG and HEATMAG
POCT Fluorescent Immunoassay Analyzer
FIA Go
New
Gold Member
Fully Automated Cell Density/Viability Analyzer
BioProfile FAST CDV

Print article

Channels

Clinical Chemistry

view channel
Image: The 3D printed miniature ionizer is a key component of a mass spectrometer (Photo courtesy of MIT)

3D Printed Point-Of-Care Mass Spectrometer Outperforms State-Of-The-Art Models

Mass spectrometry is a precise technique for identifying the chemical components of a sample and has significant potential for monitoring chronic illness health states, such as measuring hormone levels... Read more

Immunology

view channel
Image: The groundbreaking treatment approach has shown promise in hard-to-treat cancers (Photo courtesy of 123RF)

Genetic Testing Combined With Personalized Drug Screening On Tumor Samples to Revolutionize Cancer Treatment

Cancer treatment typically adheres to a standard of care—established, statistically validated regimens that are effective for the majority of patients. However, the disease’s inherent variability means... Read more

Microbiology

view channel
Image: Microscope image showing human colorectal cancer tumor with Fusobacterium nucleatum stained in a red-purple color (Photo courtesy of Fred Hutch Cancer Center)

Mouth Bacteria Test Could Predict Colon Cancer Progression

Colon cancer, a relatively common but challenging disease to diagnose, requires confirmation through a colonoscopy or surgery. Recently, there has been a worrying increase in colon cancer rates among younger... Read more

Pathology

view channel
Image: Fingertip blood sample collection on the Babson Handwarmer (Photo courtesy of Babson Diagnostics)

Unique Hand-Warming Technology Supports High-Quality Fingertip Blood Sample Collection

Warming the hand is an effective way to facilitate blood collection from a fingertip, yet off-the-shelf solutions often do not fulfill laboratory requirements. Now, a unique hand-warming technology has... Read more
Copyright © 2000-2024 Globetech Media. All rights reserved.