Leucocytes Identified by Automated Digital Morphology System

By Labmedica International staff writers
Posted on 31 Oct 2013
Image: CellaVision DM96 Digital Cell Morphology System (Photo courtesy of CellaVision).
Image: CellaVision DM96 Digital Cell Morphology System (Photo courtesy of CellaVision).
An automated digital cell morphology analyzer for determining leukocyte differential counts in peripheral blood smears (PBS) has been evaluated in a clinical laboratory.

The differential counting of peripheral blood leukocytes is one of the most frequently ordered laboratory tests by clinicians as it is an important test for the diagnosis of various hematologic diseases and systemic diseases. Improvements in laboratory technologies have benefited from the development of automated blood cell counters that have begun to replace the manual microscopic counting that is still performed in most laboratories.

Laboratory scientists at the University of Ulsan College of Medicine and Asan Medical Center (Seoul, Republic of Korea) examined a total of 308 peripheral blood samples with quantitative or qualitative abnormalities, for which complete blood counts had been ordered by clinicians. These samples were initially analyzed using the automatic blood cell analyzer Sysmex XE-2100 (Sysmex; Kobe, Japan), and the total leukocyte and differential counts were determined. Manual microscopic differential counts of 100 cells on each slide were separately performed by two independent well-trained laboratory technologists.

The slides were labeled and loaded into the CellaVision DM96 system (CellaVision AB; Lund, Sweden) after manual microscopic differential counting. In cases with a buffy coat preparation, the slides were prepared using samples prior to buffy coat preparation and labeled in the CellaVision DM96 system To evaluate the clinical relevance of the extension of cell counts up to 300 or 500 cells, which is the provided function by the CellaVision DM96 system, in the samples with low leukocyte count of less than 1,000 cells/μL, correlation analysis between the CellaVision DM96 system and manual count was performed.

The correlation coefficients between two methods were consistently high, ranged from 0.864 to 0.992. The sensitivity, specificity, positive predictive value, negative predictive values of this system for the identification of abnormalities was consistently high, especially for blast cells. When the instrument was instructed to count 300 or 500 cells from the operator, better performance was demonstrated than 100 cells in the leukopenic samples by sacrificing only 40 seconds/slide on average.

The authors concluded that the CellaVision DM96 system is useful in the clinical laboratory providing comparative accuracy compared with manual counts in samples with abnormalities. In leukopenic samples, report quality can be improved by ordering to count 300 or 500 cells from the operator without severe prolongation of turnaround time. The study was published on the October 2013 issue of the International Journal of Laboratory Hematology .

Related Links:

University of Ulsan College of Medicine and Asan Medical Center
Sysmex 
CellaVision AB 



Channels

Clinical Chemistry

view channel
Image: The ICON 25 hCG point-of-care pregnancy test (Photo courtesy of Beckman Coulter).

Pregnancy Tests Evaluated During Early Gravidity

The predominant hormone human chorionic gonadotropin (hCG) variant in urine, hCG β core fragment (hCGβcf), has been demonstrated to cause false-negative results in qualitative point-of-care (POC) hCG devices.... Read more

Genetic Tests

view channel
Image: The Quant-iT PicoGreen dsDNA Assay Kit (Photo courtesy of Invitrogen).

Blood Test Provides Rapid Method of Detecting Solid Cancers

Circulating tumor DNA (ctDNA) is a promising biomarker for noninvasive assessment of cancer burden, but existing detection methods have insufficient sensitivity or patient coverage for broad clinical applicability.... Read more

Pathology

view channel
Image: Micrograph showing prostatic acinar adenocarcinoma, the most common form of prostate cancer (Photo courtesy of Zephron).

New Test Could Accurately Predict Prostate Cancer Recurrence

A test based on genetic signature and tumor oxygen levels, which boost the test’s accuracy, can predict whether a man is at high risk of prostate cancer recurrence. Prostate cancer is the most common... Read more

Industry News

view channel

Roche and Hitachi Renew Alliance in Diagnostics

Roche (Basel, Switzerland) and Hitachi (Tokyo, Japan) have renewed their alliance in the diagnostics business by 10-years for the joint development and manufacture of the next generation of instruments and workflow automation solutions for medical laboratories. The renewed alliance marks a significant milestone towards... Read more
 
Copyright © 2000-2014 Globetech Media. All rights reserved.